Dragonfly Inc. 
Model HPD-1700

INSTALLATION AND SETUP MANUAL

 

Theory of Operation

Safety Precautions
Connection, Main View, End View

Installation 

Mounting to lab bench

Height Adjustment

Hydraulic System

Piston & Manifold Assembly

Maintenance of components 

Filling the fluid chamber 

Pressure Transducer
Piston Spacer

Charge Amplifier 

Oscilloscope Connections

Triggering Device
Timing Adjustment  Trigger Connection

Chronic Implant Luer Lock Fittings

Static Pressure Testing
Pendulum Weight  Inclinometer

Research and Abstracts

 

(Use 'back' button to return to index)

  

DRAGONFLY

 RESEARCH & DEVELOPMENT, INC.

P O BOX 507                 RIDGELEY,  WV   26753-0507                 (304) 738-3609

 

 

 

 

 

MADE IN USA

Ó  1996-2001 DRAGONFLY, R&D, INC.

 

 

  

THEORY OF OPERATION

           

The DRAGONFLY R&D INC. MODEL # HPD-1700 FLUID PERCUSSION DEVICE  is used to create reproducible closed CNS/brain trauma injury in the laboratory rat. This device consists of a closed hydraulic system actuated by a falling pendulum. The pressure with which the pendulum strikes the closed hydraulic system is adjusted by changing the angle through which the pendulum falls or by adding optional weights to the pendulum.  The system pressure is measured directly by a precisely calibrated pressure transducer through a charge amplifier/coupler to a digital or analog oscilloscope (digital dual- trace with RS-232 output is ideal), or  most PC  based data acquisition systems. The timebase of the oscilloscope in combination with the mechanically adjustable remote oscilloscope trigger allows the pressure signal to be seen at the correct time. 

            The oscilloscope must be set to trigger on a positive or rising trigger pulse. The storage function of oscilloscope should be turned on. The time duration of the impact or pressurization event occurs in approximately 10 ms with the flat faced striker plate, however, it can be lengthened to 20 ms or more by gluing shock absorbing closed cell foam on the striker plate of the piston. The time duration is read directly on the oscilloscope using screen grid as reference.

            The closed hydraulic system is attached to the animal model using a flexible high pressure hose attached to a chronically implanted Luer lock fitting on the rat's skull.  See “Surgical Attachment of Chronic Implant.”


READ THESE

SAFETY  PRECAUTIONS

BEFORE PROCEEDING!

 

·        READ ENTIRE INSTALLATION AND SETUP MANUAL BEFORE ATTEMPTING TO INSTALL OR OPERATE THIS EQUIPMENT.
      

·        WEAR EYE PROTECTION WHEN INSTALLING OR OPERATING THIS EQUIPMENT!

         

·        BEWARE OF HAND PINCH  HAZARD WHEN ADJUSTING PENDULUM ANGLE.  ALWAYS ADJUST PENDULUM ANGLE USING THE BRASS PULL-PIN.  DO NOT PUT FINGERS IN AREA BENEATH PENDULUM HEIGHT ADJUSTMENT ARM!

         

·        CAUTION:  THIS DEVICE DEVELOPS MOMENTARY HYDRAULIC PRESSURE  UP TO 200 PSI.  PROTECT YOUR EYES!

         

·        USE CAUTION AGAINST ELECTRICAL SHOCK WHEN USING LIQUIDS IN THE PROXIMITY OF  HIGH VOLTAGE ELECTRICAL EQUIPMENT SUCH AS OSCILLOSCOPES OR PC BASED DATA SYSTEMS.

         

·        KEEP HANDS CLEAR OF PENDULUM SWING WHEN PENDULUM IS IN THE STRIKE POSITION.

         

·        ALWAYS REMEMBER:       SAFETY FIRST!

  

 

INSTALLATION        

 

            The DRAGONFLY HPD-1700 is fully assembled and tested by DRAGONFLY R&D, INC.. Note that the pendulum and the pendulum vertical support assembly has been disassembled for shipping.  Remove the percussion frame assembly from its box and remove plastic covering.  To install the pendulum and the pendulum vertical support assembly,  remove the four screws holding the shipping spacer plate mounted to the bottom of the vertical supports. Cut and discard any nylon ties holding pendulum in place.  Attach the vertical supports securely to the percussion frame baseplate using the four 1/4-20 stainless steel flat head slotted screws supplied. Note that all components are fully adjustable using the tool kit supplied.  All screws should be securely tightened.  Containers and shipping spacer plate may be reused in the event of return to the manufacturer for repairs or service.

   

  

            The HPD-1700 must be mounted securely to an appropriate laboratory bench for effective operation. The laboratory table should be heavy and solid. If the HPD-1700 is not mounted solidly, secondary shock waves may appear in oscilloscope tracings.  Drill four 1/4 inch mounting holes through the laboratory bench top using the dimensions listed in the specifications section. 

  

  

Using the actual device for a drilling guide is acceptable, however use it to spot or start the hole only, then  remove the device and continue drilling through the lab bench top.  Attach baseplate of Percussion Device to the bench top with Neopreme TM washers between the bench top and the bottom of the Delrin TM baseplate spacers.  Insert 3 of the 4 each  1/4- 20 NC baseplate screws (an assortment of lengths is included) using  the Neopreme TM mounting washer assortment in conjunction with line level included in the tool kit. Use a snugging nut, a lock washer, and a locking nut beneath the bench. Do not over-tighten screws!  When the baseplate is level in all directions, insert  an assortment of mounting washers  to fit snugly between the bench top and the base of the Percussion Device. Then insert the last  mounting screw, snug nut, lock washer, and lock nut. Facing the HPD-1700 with the Piston to your right is considered “front” of the HPD-1700.

  

   

            Attach the Mechanically Adjustable Magnetic Switch Assembly securely to the rear Vertical Pendulum Support using the 2 ea. flat headed 1/4-20 slotted screws supplied. Attach the Magnetic Switch wire to the rear vertical support  through the plastic wire clip using  the 8-32 NF  plastic screw supplied. The 8-32 NF threads into the 1/4-20 Mechanical Trigger mounting screw, which has been drilled and tapped for that purpose. It is the one nearest to the pendulum end of the device. 

  


   


    

The inclinometer/protractor supplied is used to adjust the reference angle through which the pendulum falls, and can be cross-referenced with the locking protractor as shown. Unlock the Pendulum Height Arm Locking Screw (3/4 inch wrench supplied) and adjust the angle through which the Pendulum falls as shown in the photograph,  with the Pull Pin installed. When the desired angle is atained, lock the Locking Screw.  Use caution not to pinch your fingers between the Pendulum Height Arm and the Vertical Pendulum Support.

  


             HPD-1700 Height Adjustment

            

        Finally, install piston into the piston carrier plate by sliding the front diameter of the piston into position in the Piston Carrier Plate and securing with the rear of the piston with the knurled Stainless Steel Piston Mounting  Screw supplied. Attach the Guide Key for the Piston Carrier to the Baseplate.  Secure the piston carrier plate to the baseplate of the Fluid Percussion Device using the 4 each Allen Head Cap Screws and thick Stainless Steel Washers supplied. A full set of Allen TM Wrenches is supplied in the tool kit.

 



 

THE HIGH PRESSURE  PISTON AND MANIFOLD ASSEMBLY

   

Piston/Manifold Assembly 

 

   Manifold Assembly

   

            The high pressure piston and manifold assembly is supplied completely assembled and tested. Periodically, though, the piston and manifold assemblies should be disassembled and inspected for deposited contaminates. Re-assemble using the instructions in the following sectionMaintenance Of Hydraulic System Components”. For initial setup this section can be skipped.

            The piston and cylinder are used to develop hydraulic pressurization in the closed hydraulic system. Non-lubricated high pressure seals reduce introduction of contaminates into the deionized distilled water used to fill the cylinder. It is advisable to change the water often as directed in “Filling The Hydraulic Cylinder” to avoid bacterial growth or contamination in the hydraulic system, ultimately introduced to the animal model.

             The cylinder and manifold system should be removed from the  percussion platform between use and the water should be removed from the system. The system can be dried by vigorously activating the piston in the cylinder with the animal attachment end down. The piston and manifold assembly can also be washed with mild detergents such as liquid dish washing detergent in moderately warm water using the same method described. Rinse repeatedly and thoroughly removing all detergents. 

 

CAUTION!    Harsh detergents may damage high pressure seals!

 

            The stainless steel high pressure manifold is a "T" junction manifold allowing the percussion shock wave to be hydraulically transmitted from the pressurizing cylinder to both the animal model and the pressure transducer simultaneously. This allows direct recording of the actual pressure  ((voltage from transducer) /(time )) in seconds developed at the animal model via the transducer, charge amplifier/coupler and oscilloscope.

             The animal can be attached directly to the high pressure manifold if desired, or attached using the sample plastic Pressure Monitoring Line (M/F / PVC/ 6 INCH) supplied. Order replacements directly from

:  

Dragonfly Research & Development, Inc.

PO Box 507

Ridgeley,  WV  26753-0507,  USA.    (304) 738-3609

(rated 800 psi, minimal expansion under normal conditions.)

           

 A stereotaxic table is useful in positioning the animal model. The Manifold is easily positioned and held by hand or by an ordinary three finger laboratory bar clamp with a weighted base. An optional Clamp with a Marble Base and an adjustable stainless steel Arm is available from Dragonfly Research & Development, Inc.


MAINTENANCE OF HYDRAULIC SYSTEM COMPONENTS

(Initially assembly by Dragonfly, Inc.)

 

            The percussion is measured directly using a precisely calibrated piezo. Assemble the pressure transducer into pressure transducer housing provided using appropriate seals included. Do not over tighten as damage may occur to transducer! Next, assemble the transducer and housing into the 1/8 NPT thread in the end of the stainless steel high pressure manifold.  The pipe thread of the transducer housing should be wrapped with Teflon thread sealing tape.  Wrap one thickness only in counterclockwise direction, while pulling tape to conform with the pipe thread. Do not over tighten the transducer and transducer housing into the high pressure manifold.

  

 

             Wrap both ends of the high pressure hydraulic hose with Teflon pipe tape as described above.  Insert one end of the high pressure hydraulic hose into the rear fitting of the pressurizing cylinder, and the other end into the pipe thread fitting on the bottom of the stainless steel high pressure manifold. Tighten the threads so that the high pressure manifold is vertical  (i.e. perpendicular plane to table top). Do not over tighten!

            After assembly, the high pressure system can be checked for leakage using air pressure.  Fill the pressurizing cylinder with air, then block the flow of the male Luer fitting on the high pressure manifold.  Pressurize the piston by hand while submerging the entire pressurizing system,  except for the pressure transducer output connector,  under water.  If air bubbles are observed at any fitting, this is an indication of leakage, and that fitting must be corrected.  An alternative to submerging the pressurizing system is to use a solution of  mild dish washing liquid (high sudsing agent) and water, or a commercial sudsing agent such as SNOOP TM to check each fitting for leakage while pressurizing the cylinder by hand.

            If the transducer output connector gets wet, dry it with  aerosol “canned air”, or a filtered  “house” air pressure hose. Water in the Transducer  Output Connector may cause erroneous data.

 

           

FILLING THE HYDRAULIC CYLINDER

 

            To operate correctly,  the fluid cylinder must be “completely filled” with deionized, distilled water insuring that no air bubbles are trapped in the cylinder, manifold, or high pressure hoses. Trapped air bubbles will show up as momentary plateaus or flat spots on oscilloscope tracings.

             The filling procedure is very similar in theory to that used to fill a hypodermic syringe and remove any trapped air bubbles.  Filling is accomplished by pushing the piston in and out of the cylinder by hand while the high pressure hose is inserted into a container of deionized, distilled water with the stopcock in the open position (i.e. in line with the flow).  The cylinder must be physically below the water level in the container so that any trapped air bubbles can escape.  Open the stopcock on the high pressure manifold allowing water to be pulled into and out of the piston.  Operate the piston slowly by hand,  back and fourth in the cylinder until no air bubbles are observed in the exhaust stroke.. Tapping gently on cylinder, manifold, and stopcock with wooden or plastic block may help dislodge the bubbles. The piston is now filled and purged of air bubbles

             Fill the cylinder to approximately half of its stroke by slowly pushing on the piston.  The gold Anodized Piston Spacer can now be used to obtain a reproducible volume of fluid in the piston.  Close the stopcock before removing the manifold from the water container. The hydraulic cylinder is now ready to use. Install the cylinder back into position in the movable cylinder carrier of the HPD-1700, and remove the piston spacer.

  

Piston Spacer

 

             Fluid displacement occurs momentarily during percussion into and out of the animal skull due to the compressibility of brain tissues.  The minimum diameter of the closed hydraulic system is the Luer lock which is .087 inch diameter.  This figure can be used to calculate the minimum cross-sectional area of the fluid path when considering  the fluid dynamics of percussion.


 

PRESSURE TRANSDUCER CONNECTIONS

 

            The Quartz pressure transducer converts hydraulic pressure in the closed hydraulic system into voltage which is displayed as the vertical component of the oscilloscope. After mounting the pressure transducer into the manifold as described above in  “ASSEMBLY OF HYDRAULIC SYSTEM “, simply screw the transducer cable connector onto the transducer, then connect the BNC connector on the opposite end of the transducer cable into the BNC input of the charge amplifier/coupler. The transducer output is excited and AC coupled to the oscilloscope via the amplifier/coupler with less than 100 ohms output impedance.

            The maximum full scale pressure of the transducer is 200 psi. Do NOT exceed full scale pressure as permanent damage will occur. Extremely fast transients (< 2 microsecond rise time) may show transducer resonance superimposition. Sharp wave front pressures may excite passage resonance which show up as secondary transient. Resonant harmonics oscillations may also show up as secondary transients.  Large secondary transients suggest improper mounting of Fluid Percussion Frame.  Experiment using the multiple sets of rubber shock absorbers supplied for best results in reducing secondary transients. Entrapped air bubbles usually show up as plateaus on both the rising and falling waveform.

.

LINEAR CONSTANT OF TRANSDUCER SENSITIVITY = 30.76mV/psi

 

**** CAUTION: DO NOT EXCEED 200 PSI  !!! ****

 

THE CHARGE AMPLIFIER

 

            The charge amplifier/coupler requires  9 volt transistor batteries for operation. Slide the battery cover on the bottom of the charge amplifier/coupler in the direction indicated to gain access to battery connections. Install batteries. Test batteries by momentarily switching on/off switch to Battery Test position. Replace the battery cover. Replace batteries when necessary to prevent unreliable data.

            The output signal from the Pressure Transducer is attached to the BNC input of the charge amplifier/coupler.  The BNC output of the charge amplifier/coupler is connected to the AC voltage input of one channel of the Oscilloscope.

 

OSCILLOSCOPE CONNECTIONS

 

            Connect the BNC to BNC cable from the output of the charge amplifier/coupler into the  BNC oscilloscope input of channel one of the vertical deflection unit.  Connect the second BNC to BNC cable provided into the BNC connector of the remote oscilloscope trigger output and the opposite end into the REMOTE TRIGGER of the oscilloscope.   The remote trigger of the oscilloscope is generalized as the EXTERNAL triggering input on the horizontal deflection unit ( i.e. the Time Base)

 

 

OSCILLOSCOPE SETUP

 

The vertical deflection unit, ( Voltage ), should be set to 1 V/div with AC coupling on channel one, (for multi- channel oscilloscopes).  Trace alignment should be set to one major division above bottom of display, in order to facilitate maximum viewing area for waveform. Turn waveform storage function on. The horizontal deflection unit, ( Time Base), initially should be set to 10 ms/div, with no additional Delay Time factor.              Triggering is set to single sweep operation on a positive dc voltage  provided by the dc voltage amplifier of the magnetically switched remote timing unit.  Set the oscilloscope to “external trigger” using the    external timing input connector.  Triggering level initially should be set for a 50% threshold with exact value determined through trial and error.  Triggering level is set by moving pendulum  back and forth through its sweep past the magnetic trigger switch mounted on the pendulum.  Proper operation is obtained when forward pendulum movement past the  sensor triggers the  sweep of  the oscilloscope trace to occur.  Consult your oscilloscope manual for  details on “external triggering.”  If a longer timing duration is needed, the pendulum arm can be rotated 180 degrees, however, this is rarely necessary, and not advised.

 

            Sample Voltage waveform to pressure calculation:

 

            To obtain actual pressure induced in system, divide voltage by the "constant of transducer sensitivity" of 30.76mV or 0.030.76 Volts per Psig (EXAMPLE VALUE ONLY- SEE YOUR CALIBRATION SHEET FOR ACTUAL TRANSDUCER CONSTANT).

 

EXAMPLE ONLY!

 

2.6 Volts / 0.03076 V/Psig = 84.52 Psig induced.

 

Conversion of Psig to N/m squared:

1Psig = 68.95 x 1.00E -03 bar = 6894.76 N/m squared

 

THE MECHANICALLY ADJUSTABLE REMOTE  TRIGGERING DEVICE

 


Trigger Generator

    

 
Connection to Remote Triggering Device

    

            The Remote Triggering Device is activated when the magnet mounted in the pendulum swings past the magnetic switch mounted on the mechanically adjustable switch plate.  Install a 9v. battery into the "Remote Triggering Device”. Plug the  1/8 inch male connector of the magnetic switch cable into the 1/8 inch female panel mounted connector of the Remote Triggering Device”.  Plug one end of BNC cable into Remote Triggering Device” output, and the other end into the external triggering BNC input of oscilloscope.   By hand, swing the pendulum back and fourth past the magnetic trigger switch. Increase the dc voltage (+) sensitivity of the oscilloscope’s  time base external triggering function until a trace is observed, with oscilloscope set on single sweep mode. This indicates that the “Remote Triggering Device” is operating properly.

  

Trigger Timing Adjustment Assembly

 

             The closed hydraulic system must be completely assembled and filled before trigger adjustments are performed. With the stopcock/petcock closed, adjust the mechanical trigger advance/retard mechanism into its zero position. Initially adjust and lock the pendulum height adjustment arm so that the pendulum rests at approximately 20 degrees from vertical with no additional pendulum weights.  Lock the pendulum into position with the pendulum release pin. 

 

  

            Pressurize the piston by pulling the release pin, allowing the pendulum to strike the piston.  The piston must be caught by hand before a second strike occurs, otherwise repressurization will occur for each strike.  Use the time base delay function and remote trigger function of the oscilloscope in conjunction with the mechanical adjustment of the magnetic switch to locate the event occurrence on the oscilloscope screen, repeating the striking procedure as necessary until the event is visible. Moving the mechanical trigger adjustment away from the piston retards the timing of the trigger signal.  Moving the mechanical trigger adjustment toward the piston advances the timing of the trigger signal.

 

 

 

 

SURGICAL ATTACHMENT OF CHRONIC IMPLANT

 

 

            The physical fluid connection to the anesthetized animal is achieved by chronic implantation of a Luer lock fitting or “port” mounted to the animal’s skull using dental acrylic above a hole drilled through the animal skull in the area of interest. Cranial screws (OO-90 or 000-90 thread) inserted around the fitting, imbedded in the acrylic are sometimes useful.  The hole is drilled using a trephine, or a high-speed burr.

 

 

Luer Lock Fittings

  

             The surgical chronic implantation of the Luer lock fitting is of vital importance to reliable pressure readings. The minimum cross sectional diameter of the closed hydraulic system is .078 inch. Any leakage around the chronic implant will produce a lower than actual (i.e. false) pressure reading. A number of commercially available fittings have been enclosed to get you started, however, no standard industry-wide fitting currently exists. For Customized Fittings Contact:

 

Call or FAX for quotation:

Dragonfly R&D, Inc.

PH (304) 738-3609

FAX (304) 738-3607

 

            Many investigators make their own chronically implanted male Luer fittings  (headmounts) by cutting off the Luer fitting from  5 cc plastic disposable hypodermic syringes. We will produce customized prototype connections and fittings to your design and specifications. See accompanying photos of head mounted Leur ports.  Both Plastic and 316  Medical Grade Stainless Steel ports are available in a number of styles. Some models use 00-90 or 000-90 cranial screws for positive leak-free attachment.

 

 

 


STATIC PRESSURE TESTING

 

            The correct static pressure must be set before pressurizing the animal model. The static pressure (obtained by a pendulum strike with the stopcock closed) of the closed hydraulic system is set up by trial and error, starting low and working up to the desired pressure. Percussion pressure is adjusted by changing the fall angle of pendulum, or by adding or removing  pendulum weights. 

 

Pendulum Weight Assembly

 

            Use the inclinometer provided to check the angle of the pendulum for both strike position and pendulum angle. The inclinometer can be hand held. or attached to the trailing edge of the pendulum arm using a strong rubber band as is shown in the photograph.  Remove inclinometer before striking the piston.

  

Inclinometer

  

             The pressure is read directly as voltage on the oscilloscope, then divided by the linear constant of the pressure transducer (see pressure transducer calibration certificate) to calculate the effective pressure in psig. The pendulum must always be setup to strike the piston when it is perpendicular to the baseplate.  Adjust the piston into striking position by adjusting the linear position of the piston carrier plate, or by adding or subtracting water from the cylinder. Using the 3-way stopcock and water reservoir pictured makes the latter method more convenient, especially when multiple animal models are to be used ( See Refill Setup Picture).  Using a spacer of desired length to return the piston to its’ same relative position between subsequent animal models works well when using 3-way valve method with fluid reservoir (spacer included). The piston carrier plate must always be locked securely into position using the 1/4-20 NC Allen head cap screws and thick stainless steel washers , or movement of the carrier plate will occur.

 

CAUTION: DO NOT EXCEED 200 PSI.

 

            When the desired static pressure has been obtained, the stopcock is opened allowing the animal model to be subjected to the shock wave.  The actual percussion pressure (read directly as voltage on the oscilloscope) will be slightly less than the static setting due to the displacement of tissue by the deionized, distilled water entering the cranial cavity of the rat. The difference between static and dynamic pressure should be minimal since only a small volume of water is displaced.


 

DYNAMIC  FLUID PERCUSSION

 

During the static pressure test the closed hydraulic system is technically that: a closed hydraulic system. Once the stopcock/petcock is opened to the animal model the system becomes  a semi-closed hydraulic system due to the compressibility of brain tissues.

            The severity of the injury desired is obtained by observation and experimentation, using both behavioral and pathological data to grade injury.  Conditioning animal model by learned routines such as time-trial balance beams, Rota-Rods TM, grip strength measurements. treadmill, inclined plane, maze or water-maze navigation, etc. are useful in evaluating and grading the deficit incurred to the animal model  from percussion. Pathological data such as histograms may  also be used to document results.

            Record all relevant settings of  HPD-1700 for each animal model experiment,  including 1) the angular setting of the pendulum (use inclinometer or angular pointer on pivot pin of pendulum), 2) the linear setting of the piston carrier plate, 3) the linear setting of the adjustable trigger, 4) the impact position of the front edge of the  piston striker plate, 5) the amount of pendulum weights if any,  and , 6) the composition and thickness of  any closed cell foam shock absorbers used to lengthen the time frame of the event. Pressure and time data will be recorded on oscilloscope.

             Once the desired grade of injury is obtained, use the reference settings to recreate that same degree of injury to as many animal models desired.

 

 

 

 

 

      
RESEARCH AND ABSTRACTS

Here is a partial list of studies using the DRAGONFLY R&D INC. MODEL # HPD-1700 FLUID PERCUSSION DEVICE. Abstracts are available by request:
 

JOURNAL OF NEUROTRAUMA
Volume 17, Number 8, 2000
Mary Ann Liebert, Inc

Secondary Hypoxemia Exacerbates the Reduction of Visual Discrimination Accuracy and Neuronal Cell Density in the Dorsal Lateral Geniculate Nucleus Resulting from Fluid Percussion Injury

Richard A. Bauman, John J. Windholm, J.M. Petras, Kathleen McBride and Joseph B. Long

_____________________

 

NEURO REPORT
#
8 (395-398) 1997
Rapid Science Publishers

Hypoxia ptentiates traumatic brain injury-induced
expression of c-fos in rats

Jitendra R. Dave, Richard A. Bauman and Joseph B. Long

Division of Neurosciences, Walter Reed Army Institute of  Research, WRAMC, Bldg 40, Washington, DC 20307-5100 USA

_________________

 

JOURNAL OF NEUROTRAUMA
Volume 17, Number 2, 2000
Mary Ann Liebert, Inc

Real-Time Monitoring of Glutamate Following
Fluid Percussion Brain Injury With Hypoxia in the Rat

Yoshitaro Matsishita, Katsuji Shima,
Hiroshi Nawashiro and Kojiro Wada

_________________

JOURNAL OF NEUROTRAUMA
Volume 13, Number 3, 1996
Mary Ann Liebert, Inc

Laser-Doppler Flowmetry Mesurements of Subcortical
Blood Flow Changes after Fluid Percussion Brain Injury in Rats

Joseph B. Long, Jeffrey Gordon,
Joseph A. Bettencourt and Stephen L. Bolt

_________________